4. Remove the analytical column.
a. Open the front door of the GC.
b. Loosen the retaining nut from the injector and detector fitting on the upper interior
wall of the GC oven.
c. Remove the analytical column with its nut and ferrule from the bottom of the
injector and the detector.
5. Remove the bottom parts of the SSL injector. See
Figure 51. SSL Injector Bottom Parts Removal
a. Using the 1/2-in. wrench provided, unscrew the retaining nut with the base seal and
the silver seal from the bottom of the injector.
CAUTION Make sure that the liner does not come from the bottom of the injector. Save
the bottom parts of the SSL injector in a safe place because will be reused when you restore
the original configuration.
ATTENTION Assurez-vous que le liner ne sorte pas par le bas de l'injecteur. Conservez les
pièces inférieures de l'injecteur SSL en lieu sûr, car elles seront réutilisées pendant la
restauration de la configuration d'origine.
6. Install the adapter on the bottom the injector. See
Column Installation Requirements
Figure
51.
SSL Injector Bottom
Base Seal
Washer
Retaining Nut
Figure
52.
TRACE 1600/1610 Hardware Manual
1
Installation
49