GC Main Frame Advanced Maintenance
Replacing the Oven Heater Baffle
226
TRACE 1600/1610 Hardware Manual
2. Cool the oven, injectors and detectors to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
4. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
5. Remove the back cover.
a. Using a T20 Torxhead screwdriver, remove the four screws that secure the back cover
to the GC. See
Figure
Figure 202. Back Cover Removal
b. Lift the cover off using the cover handle. Be aware that the ground wire is attached to
the back panel.
Note Pay attention to the positioning of the ground wire plug, so it can be

reconnected in the same way it was removed.

6. Remove the analytical column.
202.
Back Cover Fixing Screws
Thermo Scientific