11
Upgrading Equipment
Upgrading a Stand Alone TRACE 1600/1610 to MS Version
594
TRACE 1600/1610 Hardware Manual
Figure 621. Left Panel for MS
b. Plug the ground wire previously removed to the left panel for MS provided.
c. Place the left panel for MS and attach the screw holding it in place.
5. Introduce the ISQ Series / TSQ Series transfer line inner tube into the oven through the
duct provided.
6. Attach the transfer line to the GC column using the proper nut and ferrule.
T
Note
he column must be conditioned before installing into the transfer line. See
"Installing the Column the First Time"
7. Close the front door of the GC.
8. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
9. Power on the GC.
on
page 53
Duct for ISQ Series and
TSQ Series Partial Cut
Thermo Scientific