Installing The Fpd Detector - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)

Installing the FPD Detector

190
TRACE 1600/1610 Hardware Manual
Materials needed to maintain a FPD detector
GC-grade methanol
Distilled water
Paper towels
Forceps or tweezers
FPD Jet (if necessary)
Interferential filter for Sulphur PN 28107000 or PN 19050785, (if necessary)
Interferential filter for Phosphorous PN 28107100 or PN 19050785, (if necessary)
Interferential filter for Tin PN 28107001 (optional), (if necessary)
FPD Ignition Glow-plug
T6 Torxhead key
T10 Torxhead screwdriver
5 mm wrench
5/16-in tube wrench
Note For maintaining or replacing any other component not listed in this section, see
Chapter 7, "Detectors Advanced Maintenance."
"Measuring the FPD Gas Flows"
"Cleaning or Replacing the FPD Mirror Metal Plug"
"Cleaning or Replacing the FPD Filter-side Heat Shields"
"Cleaning or Replacing the FPD Flame-side Heat Shields"
"Replacing the FPD Photomultiplier Tube"
Before proceeding the installation of the FPD detector assembly, make sure that the FPD
control module is properly installed into the back detector housing.
To install the FPD detector assembly on the FPD detector base body
1. Place the jet into the detector base body housing and tighten it. Ensure the jet is perfectly
vertically aligned to avoid damage.
2. Place the FPD on the detector base body, ensuring that the Aluminum ring has been
inserted in the correct position. Tighten the fixing nut by using the FPD fixing tool.
See
Figure
159.
on
page 309
on
page 311
on
page 334
Figure
159.
on
page 316
on
page 324
Thermo Scientific