To install a detector module
1. Remove the dummy module from the position where the detector module will be
installed.
a. Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew the two captive fixing screws.
c. Keeping the dummy module flap cover open, lift up the module from its seat in the
injector/detector housing. Place the dummy module on a clean surface.
d. Remove the gas block plug from the gas connections by unscrewing its fixing screw
using a T20 Torxhead screwdriver.
Figure 3.
Detector Gas Block Plug
WARNING Make sure all the four O-rings are placed into their seats on the gas
connection. See
Figure
2. Plug the detector module into the main frame.
a. Open the module flap cover.
b. Keeping the module flap cover open, place the module in its seat.
Be sure to insert the 25-pin male connector on the bottom of the module into the
25-pin female connector on the detector seat of the detector housing.
c. Use a T20 Torxhead screwdriver to tighten the three captive fixing screws evenly and
carefully without overtightening.
3. Do not install the module if the O-rings are missing.
Installing the Injector and Detector Modules
Gas Block Plug on Detector Seat
O-rings
Gas Connections Without Gas Block Plug
TRACE 1600/1610 Hardware Manual