11. Connect the cryo valve to the Aux Temperature/Cryo Module.
a. Connect the cable provided to the solenoid valve connector, and guide the cable
through the slot in the center of the back panel. See
Figure 559. Solenoid Valve Cable Connection
b. Connect the cryo solenoid valve to the 2-pin connector marked Cryo Valves - Oven
using the cable provided. See
Figure 560. Cryo Valves: Oven
c. Using the cable provided, connect the 15-pin female connector marked GC Bus on
the module to a Bus interface on the back of the GC.
d. Plug the power cable to the AC Input connector on the front of the module and to
the wall outlet. The LED marked On lights after the GC is powered on.
Note For further details regarding the installation of the Aux Temperature/Cryo
module, see the section
page
478.
e. Using the cable provided, connect the 15-pin female connector marked GC Bus on
the module to a Bus interface on the back of the GC.
Figure
560.
"Adding an Aux Temperature/Cryo Module"
TRACE 1600/1610 Hardware Manual
10
Adding Systems
Adding the Oven Cryo System