Upgrading Equipment
Upgrading a Stand Alone TRACE 1600/1610 to MS Version
592
TRACE 1600/1610 Hardware Manual
a. Remove the partial cut shaped plate from the exterior wall of the oven box to access
the insulating material. See
Figure 617. Perform the Duct for ISQ Series / TSQ Series (1)
b. Draw up the duct aligning the slot to the fixing holes on the exterior oven wall and
push the duct against the insulating material up to obtain a trace. See
Figure 618. Perform the Duct for ISQ Series / TSQ Series (2)
c. Using a knife or similar tool, gently cut the insulating material following the track.
d. Save the removed insulating material in a safe place because it can be reused.
Figure
617.
Partial Cut Shaped Plate
Figure
618.
Duct for ISQ Series / TSQ Series