Adding Modules
Adding a NPD Detector Module
472
TRACE 1600/1610 Hardware Manual
b. Keeping the module flap cover open, place the detector module in its seat. Be sure to
insert the 25-pin male connector on the bottom of the module into the 25-pin
female connector on the detector seat of the detector housing.
c. Use a T20 Torxhead screwdriver to tighten the three captive fixing screws without
overtightening.
IMPORTANT To maintain the correct alignment the screws must be tightened in turn.
Tighten each screw only a small amount before moving to the next screw. Repeat until all
are secure.
d. Close the module flap cover.
9. Install the NPD Thermionic Source Power Module.
The module should be installed into an external module housing provided on the back of
the GC. See
Figure
467.
Figure 467. External Modules Housing
10. Remove the cover of the external modules housing where installing the module. See
Figure
468.
External Modules Housing
Thermo Scientific