Thermo Scientific
2. In the main menu select the Instrument control icon. The Instrument Control menu
appears. In the Instrument Control menu, select the Front/Back/Aux L/Aux R GDI
detector icon to open the relevant sub-menu.
3. Set the parameters values as required, then return to main menu.
Note For details refer to Chapter 2 in the TRACE 1600/REACE 1610 User Guide.
Configure and set GDI Detector through the chromatography data system (CDS)
1. In the Configuration window select the Detectors tab.
2. Select the Detector Type: choose GDI.
3. Click GDI Config...; the GDI Configuration page is visualized.
4. Select the Gas Type used for the detector gases Gas 1, Gas 2, and Gas 3.
Choose one: Air, Hydrogen, Nitrogen, Helium, Argon, or Argon/Methane. Nitrogen is
the default gas.
5. Set the Full-scale flow of the restriction installed for each detector gas. Set a value in the
range 1-1000 mL/min. Default value is 50 mL/min.
6. Select the Max. detector temperature in the range from 0 °C to 450 °C.
The default temperature is 400 °C.
7. Select the ADC full-scale voltage. Choose one: 1 V, 5 V, or 10 V. The default value is 1V.
8. Open the GDI Method Page and set the required parameters.
Note For details refer to Chapter 4 in the TRACE 1600/REACE 1610 User Guide.
Note If the heater is installed but the actual Temperature read back is 0 °C, check the
integrity of the heater and the temperature probe.
9
Adding Modules
Adding a Generic Detector Interface
TRACE 1600/1610 Hardware Manual
527