Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)
118
TRACE 1600/1610 Hardware Manual
• Replacing the active carbon filters on the carrier gas line and split line
The active carbon filters must be replaced depending on the volume of solvent injected in
the time.
See
"Replacing the Carrier and Split Lines Filters"
• Replacing the body head O-rings
The internal (carrier line) and external (purge line) O-rings of the body head must be
replaced when there are leaks present.
See
"Replacing the Body Head O-rings"
Before maintaining the injector, read the following warning:
WARNING The injector fittings could be hot. Carry out all the operation at low
temperature to avoid burns. Therefore, before beginning the sequence, the injector must
be cooled to room temperature.
CAUTION When handling organic solvents, you must take precautions to avoid health
hazards.
ATTENTION Pendant la manipulation de solvants organiques, prenez les mesures
nécessaires pour éviter tout risque sanitaire.
Materials needed to maintain a HeS-S/SL
Septum
Tweezers
Glass liner
Liner seal (O-ring)
Ultrasonic cleaner
Mixture 1:1 methanol/acetone
Carrier gas line and/or split gas line active carbon filters
Body head internal O-ring
Body head external O-ring
T20 screwdriver
Note For maintaining or replacing any other component not listed in this section, see
Chapter 6, "Injectors Advanced Maintenance."
"Baking-out Contaminants from SSL, SSLBKF, HeS-S/SL, PTV, and PTVBKF
Injectors"
on
page 244
"Cleaning the HeS-S/SL Injector Body"
on
page
122.
on
page
124.
on
page 256
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1610

Table of Contents