Thermo Scientific TRACE 1600 Hardware Manual page 453

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
• Increase the gas flow turning the flow regulation screw counter-clockwise.
• Decrease the gas flow turning the flow regulation screw clockwise.
The standard operating value is in the range 0.2- 0.3 mL/min. Refer to the Nicolet iS50 FT-IR
Spectrometer Manual for details.
IMPORTANT The flow of the carrier gas depends on the inlet pressure. Set the carrier gas
flow according to the working pressure.
The components of the FTIR module are shown in
Figure 407. FTIR Make-up Module
To add a FTIR Make-up module
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
Figure
407.
Open/Close Knob
Flow Regulation Screw
ON/OFF Knob:
Turn clockwise for closing
Turn counter-clockwise for opening
Flow Regulation Screw:
Turn clockwise for decreasing
Turn counter-clockwise for increasing
Needle Valve VICI-CNV1A10S1
Bottom Fitting
Nut and Ferrules
TRACE 1600/1610 Hardware Manual
8
Installing Optional Kits
Installing a FTIR Make-up Module
421

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1610

Table of Contents