8
Installing Optional Kits

Installing the Pressure Regulator Kit for Gas Sampling Valve

Installing the Pressure Regulator Kit for Gas Sampling Valve
444
TRACE 1600/1610 Hardware Manual
Figure 439. Secondary FPD Detector Cables Connection
Excitation Voltage Cable
Restart the GC and set the working parameters
16. If other external modules are present, plug the power cable to the AC Input connector of
each module, and to the wall outlet.
17. Power on the GC.
a. Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker) to the position I.
Set the working parameters
18. Set the parameters for the primary FPD detector (back).
a. Set the flow of the gases, the temperatures, the ignition threshold, and the PMT
voltage for the primary photomultiplier tube according to the working analytical
conditions.
19. Set the parameters for the secondary FPD detector (front).
a. Set the flow of the gases and the temperatures to Off.
b. Set the PMT voltage for the secondary photomultiplier tube.
20. Turn on the Flame On parameter of the primary FPD detector.
This section provides the instructions for installing the kit P/N 19050763 for adjusting the
pressure of the carrier gas to correctly supply the Gas Sampling Valve (GSV) injector.
This kit is useful in the case your carrier line is not able to adjust the pressure of the carrier gas
to 450 kPa (65 psi; 4.5 bar) required for the correct functioning of the GSV injector.
Ignition Heating Cable
Signal Cable
Thermo Scientific