Adding Systems
Adding an Auxiliary Gas System
574
TRACE 1600/1610 Hardware Manual
IMPORTANT The maximum nominal inlet pressure for all the inputs is 1050 kPa (152 psig),
as indicated on the label under the gas inlets ports on the back of the GC. The working inlet
pressure range is from 400 kPa (58 psig) to 1050 kPa (152 psig).
7. Connect the Auxiliary Gas Module electrically.
a. Using the cable provided, connect the 15-pin female connector marked GC Bus on
the module to a Bus interface on the back of the GC.
IMPORTANT If two Auxiliary Gas modules are installed, one must be set as the Primary
and the other as Secondary using the dedicated switch on the back panel. See the
following example.
The primary module controls the aux pressures from 1 to 3, while the secondary module
controls the aux pressures from 4 to 6.
8. Replace the left/right side panel.
a. Plug the ground wire previously removed into the left/right panel.
b. Place the left/right panel and attach the screw holding it in place.
9. Open the gas supplies.
10. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
11. Power on the GC.
a. Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
Secondary
Primary