Thermo Scientific TRACE 1600 Hardware Manual page 511

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
6. A 2-pin connector marked Back Inlet for the connection of the solenoid valves for the
back PTV/PTVBKF cryogenic system.
7. A 2-pin connector marked Front Inlet for the connection of the solenoid valves for the
front PTV/PTVBKF cryogenic system.
8. A 2-pin connector marked Oven for the connection of the solenoid valves for the Oven
cryogenic system.
Four protecting fuses are present inside the module. See
Table 20. Aux Temperature/Cryo Module Protecting Fuses
Fuse
Type
F1
F2A 250V; (5 x 20 mm)
F2
F2A 250V; (5 x 20 mm)
F3
T2A 250V; (5 x 20 mm)
F4
F1.6A 250V; (5 x 20 mm)
The module features the following:
• The valve control of the Oven cryo option.
• The valve control of the front/back PTV cryo option.
• Sixty-three timed events to automatically drive up to eight external on/off solenoid valves.
The module should be installed into a free external module housing provided on the back of
the GC. See
Figure
476.
Figure 476. External Modules Housing
To add an Aux Temperature/Cryo Module
1. Put the GC in standby condition.
Adding an Aux Temperature/Cryo Module
Table
20.
Protections
Heater 1
Heater 2
24 V
24 V solenoid valves
External Modules Housing
TRACE 1600/1610 Hardware Manual
9
Adding Modules
479