Performing Detectors Routine Maintenance
Maintaining a Flame Ionization Detector (FID)
154
TRACE 1600/1610 Hardware Manual
Figure 118. FID Cables
Figure 119. FID Cables Disconnection
a. Loosen the glow-plug cable socket set screw using a T6 Torxhead screwdriver.
Carefully pull out the terminal body of the glow-plug cable from the dowel pin on
the glow-plug element.
b. Unscrew and pull out the straight plug crimp connector of the signal cable from the
collecting electrode bulkhead jack.
c. Pull out the quick coupling straight jack connector of the polarizing cable from the
polarizing electrode bulkhead jack.
d. Carefully move the cables in order to have free space for handling the detector.
9. Remove the top parts of the detector.
Signal Cable Straight Plug Crimp
Connector
Socket Set Screw
Glow-plug
Polarizing Cable Straight Jack Connector
Collecting Electrode Bulkhead Jack
Glow-plug Dowel Pin
Polarizing Electrode Bulkhead Jack