Thermo Scientific TRACE 1600 Hardware Manual page 363

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 310. Filter Support Reinstallation
11. Reassembly the filter and the photomultiplier assembly.
a. Reinsert the filter into the support. The mirror face must be oriented towards the
flame. See
Figure
Figure 311. Filter Remounting
CAUTION Avoid touching the filter with your fingers. If you see fingerprints on the filter,
clean it using GC-grade methanol and air dry before remounting.
ATTENTION Évitez de toucher le filtre avec vos doigts. Si vous voyez des traces de doigts sur
le filtre, nettoyez-le avec du méthanol de qualité GC et séchez-le à l'air avant de le remonter.
311.
7
Detectors Advanced Maintenance
Cleaning or Replacing the FPD Flame-side Heat Shields
TRACE 1600/1610 Hardware Manual
331