Thermo Scientific TRACE 1600 Hardware Manual page 402

Gas chromatographs
Table of Contents

Advertisement

8
Installing Optional Kits
Installing the HS Adapter Kit on the SSL/SSLBKF Injector
370
TRACE 1600/1610 Hardware Manual
14. Screw and finger-tighten the septum cap provided. See
Figure 354. HS Adapter (6)
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
15. Insert the transfer line needle into the septum cap. See
Figure 355. Transfer Line Connection
16. Cut the piece of column end previously connected to the injector.
Figure 354
Figure
355.
Thermo Scientific