7
Detectors Advanced Maintenance
Cleaning or Replacing the FID Jet
290
TRACE 1600/1610 Hardware Manual
WARNING The collector insulator must be stored somewhere it can stay clean.
b. Loosen the polarizing electrode fixing screw using a T10 Torxhead screwdriver, and
extract the polarizing electrode from its seat. See
Figure 253. FID Jet (2)
c. Place the polarizing electrode on a clean surface.
10. Remove, clean, and reinstall the jet.
a. Use the elbowed box wrench for unscrewing and extracting the jet through the
detector body. See
Figure 254. FID Jet (2)
Figure
254.
Figure
253.
Jet
Polarizing Electrode
Fixing Screw
Polarizing Electrode
Jet
Thermo Scientific