Adding Systems
Adding the Oven Cryo System
542
TRACE 1600/1610 Hardware Manual
c. Fix the solenoid valve assembly to the back of the GC using the two fixing screws
previously removed. See
Figure 556. Solenoid Valve Assemble Fixing
10. Connect the cryogenic tank tube to the solenoid valve assembly.
a. Connect the proper end of the cryo supply tube to the 1/8-in. NPT connection of
the solenoid valve using the proper nut and ferrule. Use a 7/16-in. wrench for
tightening the fittings. See
Figure
556.
Figure 557
and
Figure
Solenoid Valve
Assembly
Fixing Screws
558.