Installation
Coupling to a Mass Spectrometer
26
TRACE 1600/1610 Hardware Manual
Figure 29. Pre-shaped Plug Removal
IMPORTANT Save the pre-shaped plug of insulating material in a safe place because it
could be reused.
c. On the interior of the oven box remove the partial cut plate from the corresponding
duct. See
Figure
Figure 30. Duct for the Coupling with ISQ Series or TSQ Series Mass Spectrometer
d. Place the left panel for your MS and attach the screw holding it in place.
3. Introduce the transfer line inner tube into the oven through the duct provided.
4. Attach the transfer line to the GC column using the proper nut and ferrule.
CAUTION - INSTRUMENT DAMAGE: Make sure that the GC column has been
conditioned before connecting it to the transfer line. The material released from the
column, (column bleed), during conditioning may contaminate the detector.
30.
ISQ Series and TSQ Series Duct
Duct for ISQ Series /TSQ Series
Pre-shaped Plug