Thermo Scientific TRACE 1600 Hardware Manual page 611

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 603.
View of the Sensor Tube Duct Into the Oven
b. Check the duct is free of insulating material; if not, remove it.
IMPORTANT The duct must be free of the insulating material. If not, it could obstruct
the tube sensor.
10. Install the hydrogen sensor.
a. Move and guide the sensor tube into the duct for the hydrogen sensor. See
Figure
604.
Figure 604. Hydrogen Sensor Installation (1)
b. Place the hydrogen sensor into the back of the GC aligning the fixing holes to the
corresponding holes on the GC chassis. See
10
Adding the Hydrogen Sensor
Sensor Tube Duct
Fixing Holes
Sensor Tube
Figure
605.
TRACE 1600/1610 Hardware Manual
Adding Systems
579