Adding Modules
Adding a PDD Module
496
TRACE 1600/1610 Hardware Manual
.
IMPORTANT To maintain the correct alignment the screws must be tightened in turn.
Tighten each screw only a small amount before moving to the next screw. Repeat until all
are secure.
9. Mount the support bracket of the pulse generator. See
Note The support bracket can hold two pulse generators for supplying two PDD
detectors on the GC.
Figure 495. Support Bracket
a. Insert the upper clamps of the support bracket into the last slots of the cooling fan
grid. See
Figure
Figure 496. Support Bracket Installation (1)
b. Lift the bottom part of the support bracket and push it back until the lower lamp
hooks the rear of the GC top cover. See
496.
Upper Clamps
Figure
Figure
495.
Lower Clamp
Upper Clamps
497.
Thermo Scientific
Cooling Fan
Grid