Detectors Advanced Maintenance
Cleaning or Replacing the FPD Filter-side Heat Shields
318
TRACE 1600/1610 Hardware Manual
Figure 288. FPD Detector Removal
Note Do not lose the Aluminum ring inserted between the detector head and
the base body.
7. Remove the photomultiplier assembly and the filter.
a. Loosen the knurled nut that fixes the photomultiplier assembly and remove it from
the detector body. See
Figure 289. Photomultiplier Assembly Removal
Figure
289.
FPD Fixing Nut
Aluminium Ring
Knurled Nut
Thermo Scientific