GC Main Frame Advanced Maintenance
Replacing the Flap Motor
240
TRACE 1600/1610 Hardware Manual
Figure 214. Back Cover Removal
b. Lift the cover off using the cover handle. Be aware that the ground wire is attached to
the back panel.
Note Pay attention to the positioning of the ground wire plug, so it can be
reconnected in the same way it was removed.
6. Remove the AI/AS autosampler if present.
7. Remove the right panel.
a. Open the front door of the GC.
b. Using a T20 Torxhead screwdriver remove the screw located on the upper right
corner of the frontal frame.
c. Slide the panel towards the back of the instrument. Tilt the top of the right panel
outwards.
d. Continue to slide the panel towards the back of the GC and lift it off.
8. Remove the cooling duct. See
Figure
215.
Back Cover Fixing Screws