Figure 65. Inserting the Column into the iConnect
10. Insert the column through the iConnect nut and ferrule (open end up).
11. Set the column length according to the injector or detector type. See
Table 12
for column insertion depths.
Note Detector and injector insertion depths are measured from the top of the ferrule.
See
Figure 66
for an example of the SSL and SSLBKF insertion depths.
1
iConnect Column Lock
Push the spring to slide the
column through the iConnect
Table 11
and
TRACE 1600/1610 Hardware Manual
Installation
63