Installing Optional Kits
Installing the NoVent Microfluidics
412
TRACE 1600/1610 Hardware Manual
13. Place a septum with a notch cut into it behind the transfer line nut. The septum marks
the place on the column where it should exit the nut.
Figure 394. Positioning the Septum
Column
Measuring Tool
14. Pull the column back from the transfer line nut. Do not move the septum from its
position on the column.
Figure 395. Pulling the Column Back from the Transfer Line Nut
Column
Measuring Tool
Transfer Line
Nut
15. Loosen the transfer line nut from the column measuring tool.
16. Remove the column, transfer line nut and ferrule from the column measuring tool,
making sure not to move the septum from its location on the column.
17. Insert the column into the transfer line.
Transfer Line
Nut
Column
Septum
Septum
Thermo Scientific