Thermo Scientific TRACE 1600 Hardware Manual page 554

Gas chromatographs
Table of Contents

Advertisement

9
Adding Modules
Adding a Generic Detector Interface
522
TRACE 1600/1610 Hardware Manual
IMPORTANT The inlet ports of the manifold on the GDI mechanical module are
numbered 1, 2, and 3 respectively. Pay attention to the correct order when you connect
each detector gas tubing to the corresponding outlet ports marked Gas 1, Gas 2, and
Gas 3 on the GDI electrical interface.
Connect the detector gas tubing for the third-party detector as follows:
• Outlet port Gas 1 to inlet port 1
• Outlet port Gas 2 to inlet port 2
• Outlet port Gas 3 = Air (Wall) to inlet port 3
b. Repeat
step a
until all the required detector gas tubing are connected to the GDI
mechanical module.
c. Bend and run the detector gas tubing along the top cover.
Note The bending of the detector gas tubing shown in
indicative.
Figure 530. Bending of the Detector Gas Tubing on the TRACE 1600/1610 GC
Figure 530
and
Figure 531
Third-party
Detector
is
Gas Tubing
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1610