Figure 179. Glow-plug Replacing
a. Loosen the glow-plug cable socket set screw using a T6 Torxhead screwdriver.
Carefully pull out the terminal body of the glow-plug cable from the dowel pin on
the glow-plug element.
b. Using an 8-mm wrench, unscrew and remove the defective glow-plug with its washer.
c. Replace the glow-plug with a new one interposing the new washer.
9. Reinstall the left and right covers of the FPD detector body with the insulating material
in the original position, and fix them with the four fixing screws. See
CAUTION When removing the covers, pay attention to the heater element and the
temperature probe.
ATTENTION Pendant la dépose / réinstallation des couvercles, faites attention à l'élément
chauffant et à la sonde de température.
Figure 180. Detector Body Remounting
4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
Glow-plug
Glow-plug Dowel Pin
Socket Set Screw
Heater Element
temperature probe
Top Screws
Right Cover
Front Screw
Back Screw
TRACE 1600/1610 Hardware Manual
Figure
180.
205