Tip To avoid the contact with the hot air from the vents, it is suggested to gather the
electrical cables into cables holder.
a. Using the cable provided, connect the 15-pin female connector marked GC Bus on
the module to a Bus interface on the back of the GC.
b. Connect up to four channels CH-1, CH-2, CH-3, CH-4 to your device selecting the
full scale of each analogue signal from 1 V to 10 V according to your needs.
IMPORTANT Only connect the Ground wire to the GND contact of the AOI module or
of the your device. Do not connect the Ground cable to both the devices.
c. Connect the Start OUT contact closure to Start IN TTL line of the Data system
analog acquisition box.
4. Open the gas supplies.
5. If other external modules are present, plug the power cable to the AC Input connector of
each external module, and to the wall outlet.
6. Power on the GC.
a. Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
7. Configure and enable the system installed through the touch screen of your GC. Refer to
the TRACE 1600/1610 User Guide.
9
Adding Modules
Adding an Analog Output Interface
TRACE 1600/1610 Hardware Manual
531