Figure 551. Aux Temperature/Cryo Module Installed into the GC
9. Install the solenoid valve.
The solenoid valve must be installed in the proper seat on the back of the GC. See
Figure
552.
Figure 552. Solenoid Valve Assembly Seat
a. With care, take the solenoid valve assembly and remove the two screws from the
brackets using a T15 Torxhead screwdriver. These screws are used to fix the bracket
on the GC.
If not already done, connect the coolant tube to the solenoid valve using the proper
nut and ferrule. Use a 7/16-in. wrench for tightening the fittings. See
Cables Holder
Aux Temperature/Cryo Module
TRACE 1600/1610 Hardware Manual
10
Adding Systems
Adding the Oven Cryo System
Figure
553.
539