Figure 471. NPD Thermionic Source Power Module Installed into the GC
12. Connect the NPD Thermionic Source Power Module.
Tip To avoid the contact with the hot air from the vents, it is suggested to gather the
electrical cables into cables holder.
a. Open the module flap cover.
b. Using the cable provided, connect the source assembly connector on the NPD
Thermionic Source Power Module to the thermionic source and twist the ring to lock
the connection. See
Figure 472. Thermionic Source Assembly Cable Connection
Cables Holder
NPD Thermionic Source Power Module
Figure
472.
9
Adding Modules
Adding a NPD Detector Module
TRACE 1600/1610 Hardware Manual
475