Performing the Dual FPD Detector Configuration

Thermo Scientific
This section provides the instructions for performing the Dual FPD configuration by
connecting a second photomultiplier tube with a different interferential filter on the same
detector control module. See
Figure 422. Dual FPD Configuration
To perform the Dual FPD configuration you need the following devices:
• Two Flame Photometric Detector modules
• A Dual FPD upgrade kit PN 19050783
IMPORTANT The Dual FPD must be installed onto the FPD control module placed into
the back site of the detectors housing.
The FPD control module placed into the front site of the detector housing must be
equipped with the PT100 simulator, provided with the kit, instead of the FPD detector,
and will supply the second photomultiplier of the configuration.
The operation consists of the following steps:
• Remotion of the front and back FPD Detectors from the relevant detector control
module.
• Disassembling the photomultiplier tube from a FPD detector.
• Assembling the second photomultiplier tube on the other FPD detector to obtain the
Dual FPD configuration.
• Remotion of the jet from the front detector control module.
• Installation of the PT100 simulator, provided with the kit, on the detector base of the
front detector control module.
Figure
422.
8
Installing Optional Kits
Performing the Dual FPD Detector Configuration
TRACE 1600/1610 Hardware Manual