Table 4. Ferrules for Injectors and Detectors
Ferrule Type
Encapsulated
graphite ferrule
See the Note
below.
Graphite ferrules are a soft material that is porous to oxygen, making them suitable for most
applications except GC/MS interface connections. These easy-to-use ferrules form a soft grip with the
column and provide a stable seal. Graphite ferrule are re-usable.
Graphite/Vespel® ferrules have a long lifetime and are compatible with GC/MS. These ferrules form
a strong grip with the column and cannot be reused as they form a permanent seal with the column.
They have a temperature limit of 350 °C, but must be re-tightened after initial temperature cycles.
Note An optional adapter kit must be installed if you want to connect the column to
the injector and detector by using the encapsulated graphite ferrule. See
the Adapters for Encapsulated Graphite Ferrules"
Table
5,
Table
6, and
Table 7
diameter of precolumn and capillary column.
Table 5. Graphite Ferrules Size ID
Ferrules Size ID
Required for:
0.4 mm
0.1 - 0.32 mm ID capillary column
0.8 mm
0.45 - 0.53 mm ID capillary column
Table 6. Graphite Vespel Ferrules Size ID
Ferrules Size ID
Required for:
0.4 mm
0.1/0.2/0.25 mm ID capillary column
0.5 mm
0.32 mm ID capillary column
0.8 mm
0.53 mm ID capillary column
Table 7. Encapsulated Graphite Ferrules Size ID
Ferrules Size ID
Required for:
0.4 mm
0.25 mm ID capillary column
0.5 mm
0.32 mm ID capillary column
0.8 mm
0.45/0.53 mm ID capillary column
Injector/Tee
Detector
Connector
PTV, PTVBKF,
SSLBKF
(Tee Connector)
on
list the size of the ferrule to use depending on the internal
1
Column Installation Requirements
Temperature
Limit
450 °C
"Installing
page
48.
TRACE 1600/1610 Hardware Manual
Installation
47