Figure 255. NPD Signal Cable Removal
8. Open the module flap cover.
9. Remove the thermionic source assembly cable. See
Figure 256. NPD Thermionic Source Cable Removal
a. Twist the ring to disconnect the thermionic source assembly cable.
b. Push and twist the lock so that the button slides up in the groove, then pull the cable
ends apart.
10. Unscrew and pull out the straight jack connector of the collecting electrode cable from
the collecting electrode bulkhead jack. See
7
Detectors Advanced Maintenance
Measuring the NPD Gas Flows
Figure
256.
Thermionic Source Assembly
Thermionic Source Assembly Cable
Figure
257.
TRACE 1600/1610 Hardware Manual