It will be necessary to push down on the right side of the module near the back edge to
ensure engagement of the 25 pin D-sub electrical connector. Wiggle the module as
necessary to engage the pins and then secure the screws.
8. At this point, the regulator that delivers the helium can be set to the appropriate pressure:
typically 110 psig (760 kPa; 7.6 bar) will suffice for 0.25 mm i.d. columns of 30 m
length.
Dial the pressure up slowly while tapping the pressure gauge in order to set the
appropriate pressure.
pressure.
Table 21. Determining Correct Helium Regulator Pressure for Column Type (Sheet 1 of 2)
Column Length
5 m/10 m
10 m/20 m
15 m
30 m
60 m
100 m
Table 21
serves as a guide for setting the correct helium regulator
Column i.d.
Desired Flow
0.10 mm
0.4 mL/min
0.5 mL/min
0.6 mL/min
0.18 mm
0.8 mL/min
1.0 mL/min
1.2 mL/min
1.5 mL/min
2.0 mL/min
0.25 mm
1.0 mL/min
1.2 mL/min
1.7 mL/min
2.2 mL/min
2.7 mL/min
0.25 mm
1.0 mL/min
1.2 mL/min
1.7 mL/min
2.2 mL/min
2.7 mL/min
3.2 mL/min
0.25 mm
1.0 mL/min
1.2 mL/min
1.7 mL/min
2.2 mL/min
0.25 mm
1.0 mL/min
9
Adding Modules
Adding a Helium Saver Injector Module
Required helium regulator
*
pressure psig (kPa; bar)
110 psig (760 kPa; 7.6 bar)
120 psig (830 kPa; 8.3 bar)
130 psig (900 kPa; 9 bar)
100 psig (690 kPa; 6.9 bar)
110 psig (760 kPa; 7.6 bar)
120 psig (830 kPa; 8.3 bar)
130 psig (900 kPa; 9 bar)
140 psig (970 kPa; 9.7 bar)
100 psig (690 kPa; 6.9 bar)
110 psig (760 kPa; 7.6 bar)
120 psig (830 kPa; 8.3 bar)
130 psig (900 kPa; 9 bar)
140 psig (970 kPa; 9.7 bar)
100 psig (690 kPa; 6.9 bar)
110 psig (760 kPa; 7.6 bar)
120 psig (830 kPa; 8.3 bar)
130 psig (900 kPa; 9 bar)
140 psig (970 kPa; 9.7 bar)
150 psig (1030 kPa; 10.3 bar)
120 psig (830 kPa; 8.3 bar)
130 psig (900 kPa; 9 bar)
140 psig (970 kPa; 9.7 bar)
150 psig (1030 kPa; 10.3 bar)
130 psig (900 kPa; 9 bar)
TRACE 1600/1610 Hardware Manual
**