a. Remove the protective cap covering the new thermionic source.
b. Mount and guide the new source assembly on the NPD body proceeding in the
reverse order in which it was removed. Be careful not to bump the bead on the sides
of the body and collecting electrode.
c. Rotate and align the thermionic source connector, then tighten the dowel to fix the
connector to the support using the T6 Torxhead key. See
Figure 276. Reinstall Thermionic Source (1)
18. Reconnect and screw the straight jack connector of the collecting electrode cable to the
collecting electrode bulkhead jack. See
Figure 277. Reinstall Thermionic Source (2)
19. Reconnect the source assemble cable to the NPD cable and twist the ring to lock the
connection. See
Figure
Figure
277.
Collecting Electrode Cable
Collecting Electrode Straight Jack Connector
278.
7
Detectors Advanced Maintenance
Cleaning or Replacing the NPD Jet
Figure
276.
Correct Alignment
Dowel
TRACE 1600/1610 Hardware Manual
307