Installing The Oven Exhaust Kit - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

8
Installing Optional Kits

Installing the Oven Exhaust Kit

Installing the Oven Exhaust Kit
346
TRACE 1600/1610 Hardware Manual
The oven vents at the back of the GC discharge hot air up to 450 °C (842 °F) during cooling.
Installing the optional Oven Exhaust Kit (PN 19050760), the hot air from the oven vents can
be carried away to a fume hood or other exhausting devices
The kit includes two air ducts and two 3-m length extensible Aluminum tubes.
See
Figure
326.
Figure 326. Air Duct and Extensible Aluminum Tube
Extensible Aluminum Tube
Air Duct
To install the oven exhaust kit
1. Put the GC in standby condition.
2. Cool the oven to room temperature.
3. Place each air duct on the oven vent inserting the hooks of the air duct into the relevant
vertical slots provided on the GC back panel.
4. Place an end of each extensible Aluminum tube on each air duct.
5. Connect the other end of the extensible Aluminum tube to an exhausting device.
6. Set the normal GC working conditions.
Extensible Aluminum Tube
Air Duct
Slot
.
Thermo Scientific