7
Detectors Advanced Maintenance
Cleaning or Replacing the FPD Flame-side Heat Shields
326
TRACE 1600/1610 Hardware Manual
Figure 300. FPD Detector Removal
Note Do not lose the Aluminum ring inserted between the detector head and
the base body.
7. Remove the photomultiplier assembly and the filter. See
Figure 301. Photomultiplier Assembly Removal
a. Loosen the knurled nut that fixes the photomultiplier assembly and remove it from
the detector body.
FPD Fixing Nut
Aluminum Ring
Figure
301.
Knurled Nut
Thermo Scientific