c. Fix the photomultiplier tube housing to the photomultiplier tube holder.
Carefully rotate the photomultiplier tube housing up to matching its fixing holes to
the fixing holes on the photomultiplier tube holder. See
Figure 322. Photomultiplier Tube Housing Reinstallation (2)
d. Insert the two fixing screws into the corresponding fixing holes, then tighten the
screws using a crosshead screwdriver.
Figure 323. Photomultiplier Tube Housing Reinstallation (3)
Fixing Screw
10. Reinstall the FPD detector on the base.
7
Detectors Advanced Maintenance
Replacing the FPD Photomultiplier Tube
Figure
322.
Fixing Screw
TRACE 1600/1610 Hardware Manual
341