Thermo Scientific TRACE 1600 Hardware Manual page 189

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
6. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
7. Open the module flap cover.
8. Disconnect the signal cable, glow-plug and polarizing cables from their contacts on the
cell top cover.
Figure 121. FID Cables
Figure 122. FID Cables Disconnection
a. Loosen the glow-plug cable socket set screw using a T6 Torxhead screwdriver.
Carefully pull out the terminal body of the glow-plug cable from the dowel pin on
the glow-plug element.
4
Performing Detectors Routine Maintenance
Maintaining a Flame Ionization Detector (FID)
Signal Cable Straight Plug Crimp
Connector
Socket Set Screw
Glow-plug Cable Terminal body
Polarizing Cable Straight Jack Connector
Collecting Electrode (Anode)
Bulkhead Jack
Glow-plug Dowel Pin
Polarizing Electrode Bulkhead Jack
TRACE 1600/1610 Hardware Manual
157