Detectors Advanced Maintenance
Cleaning or Replacing the FPD Mirror Metal Plug
312
TRACE 1600/1610 Hardware Manual
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If other external modules are present, unplug the power cable from the AC Input
connector of each module, and from the wall outlet.
6. Remove the FPD detector from the base.
a. Disconnect the signal, excitation voltage, and ignition/heating cables from the
detector. See
Figure
Figure 281. FPD Cables Disconnection
b. Using the tool provided with the system, loosen the fixing nut on the base of the
detector and remove it. See
281.
Figure
282.