Thermo Scientific TRACE 1600 Hardware Manual page 438

Gas chromatographs
Table of Contents

Advertisement

8
Installing Optional Kits
Installing the NoVent Microfluidics
406
TRACE 1600/1610 Hardware Manual
Figure 381. Pre-swaging the Ferrule to the 30 cm Fused Silica Tubing
SilTite FingerTite Pre-swage Tool
3. Insert the ferrule and the tubing into the NoVent microfluidics module. See
Figure 382. Securing the 30 cm Fused Silica Restrictor to the No-Vent Microfluidics Module
4. Add the FingerTite screw to the module and tighten.
5. Now you must prepare the other end of the tubing to connect to the microfluidics 3-port
splitter. It is much easier to attach the ferrule to the tubing while it is outside the GC.
You will connect it to the splitter later.
6. Position a SilFlow nut and SilFlow ferrule onto the tubing as shown in
Figure 383. Positioning the SilFlow Nut and SilFlow Ferrule Correctly on the Tubing
7. Use a scoring wafer to cut the tubing after inserting it through the ferrule. See
Then use the SilFlow pre-swage tool to secure the ferrule into position. When done
properly, the tubing will extend slightly past the tip of the ferrule.
Ferrule
Figure
382.
Figure
383.
Figure
384.
Thermo Scientific