Thermo Scientific TRACE 1600 Hardware Manual page 353

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
b. Insert the filter support into the flange and fix it tightening the three Allen screws.
Check the O-ring that fixes the heat shield is in place before tightening the screws.
See
Figure
294.
Figure 294. Filter-side Heat Shield Replacement (2)
10. Remount the filter and the photomultiplier assembly.
a. Reinsert the filter into the support. The mirror face must be oriented towards the
flame. See
Figure
Figure 295. Filter Remounting
CAUTION Avoid touching the filter with your fingers. If you see fingerprints on the filter,
clean it using GC-grade methanol and air dry before remounting.
ATTENTION Évitez de toucher le filtre avec vos doigts. Si vous voyez des traces de doigts
sur le filtre, nettoyez-le avec du méthanol de qualité GC et séchez-le à l'air avant de le
remonter.
295.
7
Detectors Advanced Maintenance
Cleaning or Replacing the FPD Filter-side Heat Shields
TRACE 1600/1610 Hardware Manual
321