Thermo Scientific
CAUTION If the filter installed is equipped with the spacer (see the kits PN 19050785 and
PN 19050786), first insert the filter, and then the spacer.
Interferential Filter
Spacer
ATTENTION Si le filtre installé est équipé d'un espaceur (voir les kits réf. 19050785 et
réf. 19050786), insérez d'abord le filtre, puis l'espaceur.
b. Mount the photomultiplier tube assembly then fix it tightening the knurled nut.
See
Figure
433.
Figure 433. Photomultiplier Tube Assembly Mounting
The result of the operation is shown in
Figure 434. Result of the Dual FPD Configuration
Primary Photomultiplier Tube
Installing of the Dual FPD and electrical connections
12. Remove the jet of the front detector control module.
a. Using the 5 mm wrench provided with the GC, loosen the jet and remove it from the
detector base body. See
Performing the Dual FPD Detector Configuration
Figure
434.
Figure
435.
8
Installing Optional Kits
Secondary Photomultiplier Tube
TRACE 1600/1610 Hardware Manual
441