Thermo Scientific TRACE 1600 Hardware Manual page 604

Gas chromatographs
Table of Contents

Advertisement

10
Adding Systems
Adding an Auxiliary Gas System
572
TRACE 1600/1610 Hardware Manual
Note The length of the tubes allows them to reach the Auxiliary Gas Interface whether
they are installed on the same side or on the opposite side of the Auxiliary Gas Module.
IMPORTANT The three gas tubes, coming from the gas tubing block, are numbered 1, 2,
and 3 respectively. Pay attention to the correct order when you connect each tube to the
corresponding inlet on the auxiliary gas interface. The end of each tube is provided with a
label indicating the type of gas.
b. Bend the gas tube until its end reaches the corresponding numbered inlet port of the
auxiliary interface.
c. Connect the gas tube to the corresponding numbered inlet port using the appropriate
nut and ferrules. Use a 7/16-in. wrench to tighten the fittings. See
Figure 597. Gas Tubes Connection to the Auxiliary Gas Interface - External View(
d. Repeat
step b
and
auxiliary gas interface.
e. In the GC oven, carry out the connections of the components of interest to the
corresponding inlet ports of the auxiliary gas interface using the appropriate nut and
ferrules.
step c
until all the gas tubes of interest are connected to the
Figure
597.
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1610

Table of Contents