Figure 170. Photomultiplier Assembly Removal
a. Loosen the knurled nut that fixes the photomultiplier assembly and remove it from
the detector body.
CAUTION The photomultiplier tube could damage if exposed to ambient light with the
excitation voltage On. Make sure the power supply has been switched off before
disconnecting the tube from the detector body.
ATTENTION Le tube photomultiplicateur pourrait être endommagé en cas d'exposition à
la lumière ambiante alors que la tension d'excitation est activée. Assurez-vous que
l'alimentation a été coupée avant de débrancher le tube du corps du détecteur.
b. Remove the interferential filter from its housing, handling it very gently.
Keep it using a clean paper towel. See
Figure 171. Interferential Filter Removal
CAUTION Filters are fragile. Pay attention not to let the filter fall down and damage.
ATTENTION Les filtres sont fragiles. Assurez-vous de ne pas faire tomber et endommager
le filtre.
4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
Knurled Nut
Figure
171.
TRACE 1600/1610 Hardware Manual
199