Detectors Advanced Maintenance
Cleaning or Replacing the FPD Flame-side Heat Shields
330
TRACE 1600/1610 Hardware Manual
Figure 308. Filter-side Heat Shield Replacement (2)
e. Using a clean paper towel, clean the mirror surface of the plug. If necessary, use a
solvent as methylene chloride or methanol to remove deposits and a metal polishing
paste to restore it to the previous reflectivity.
f.
Insert the mirror plug in the detector body and fix it with the relevant Allen screws.
See
Figure
309.
Note Before reinserting the plug, inspect the graphite seal inside the plug's
housing. If the seal is damaged and could not ensure tightness, remove it and
replace with a new one.
Figure 309. Mirror Plug Reinstallation
g. Insert the filter support into the flange and fix it tightening the three Allen screws.
Check the O-ring that fixes the heat shield is in place before tightening the screws.
See
Figure
310.
Graphite Seal