Thermo Scientific TRACE 1600 Hardware Manual page 514

Gas chromatographs
Table of Contents

Advertisement

9
Adding Modules
Adding an Aux Temperature/Cryo Module
482
TRACE 1600/1610 Hardware Manual
Figure 480. Aux Temperature/Cryo Module Installed into the GC
8. Connect the Aux Temperature/Cryo module.
Tip To avoid the contact with the hot air from the vents, it is suggested to gather the
electrical cables into cables holder.
a. Using the cable provided, connect the 15-pin female connector marked GC Bus on
the module to a Bus interface on the back of the GC.
If the cryo option for the Oven is present, connect the 2-pin connector marked
Cryo Valves - Oven to the cryo solenoid valve using the cable provided.
If the cryo option for the front/back PTV/PTVBKF injector is present, connect
the 2-pin connector marked Cryo Valves - Front Inlet or Cryo Valves-Back
Inlet to the cryo solenoid valve using the cable provided.
If the on/off activation of external solenoid valves (up to eight) is required,
connect a 2-pin connector marked Event 1÷8 to each external solenoid valve.
Plug the power cable to the AC Input connector on the front of the module and
to the wall outlet. The LED marked On lights after the GC is powered on.
IMPORTANT The Auxiliary Temperature/Cryo module must be powered at the same
power supply voltage required for the GC.
9. Open the gas supplies.
10. If other external modules are present, plug the power cable to the AC Input connector of
each external module, and to the wall outlet.
Cables Holder
Aux Temperature/Cryo Module
Thermo Scientific