Connecting a SSL/PTV Backflush System for High Temperature

Connecting a SSL/PTV Backflush System for High Temperature
390
TRACE 1600/1610 Hardware Manual
This section provides instructions for connecting your SSLBKF/PTVBKF injector modules
with precolumn and analytical capillary column into the oven for application at high
temperature. For the purpose, the kit PN 19050254 is required. Refer to the chapters
SSLBKF Injector Module and PTVBKF Injector Module in the TRACE 1600/1610 User
Guide. This system operates at a temperature more than 300 °C.
The operation consists of the following operating procedures:
"To install the mounting bracket"
"To connect backflush line, precolumn, and capillary column"
"To install of a second backflush system"
CAUTION Before starting make sure the SSLBKF/PTVBKF injector module is correctly
installed into its seat.
ATTENTION Avant de commencer, vérifiez que le module d'injecteur SSLBKF / PTVBKF
est bien installé dans son logement.
To install the mounting bracket
Install the mounting bracket near the front of the GC oven on the right-hand side. This will
keep the mounting bracket out of the way of the column.
Proceed as follows:
1. Remove the nut at the top end of the mounting bracket. This nut secures the two sections
of the mounting bracket. See
Figure 367. Mounting Bracket Parts
Bottom
2.Push the top of the mounting bracket down.
3. Line up the bottom of the mounting bracket into two holes on the bottom side of the GC
oven.
4. Hold the mounting bracket directly upright, and twist the top end of it so that it is
securely attached to the GC oven holes.
on
page 390
on
page 396
Figure
367.
Tee Connector
on
page 392
Nut Top