Installing A Gdi Mechanical Module - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

9
Adding Modules
Adding a Generic Detector Interface

Installing a GDI Mechanical Module

518
TRACE 1600/1610 Hardware Manual
Figure 524. VOBO-HRM Board Layout
J29
J29
b. Connect the detector cable to the connector J29 AUX DET LEFT, or J13 AUX
DET RIGHT accordingly.
4. Re-couple the TRACE 1600 Auxiliary Oven to the GC.
Note Continue the procedure of installation following the instruction reported in the
section
"Installing a GDI Mechanical Module"
This section provides instructions for adding a Front/Back/Aux L/Aux R GDI mechanical
module. According to the configuration of your TRACE 1600/1610 GC, a dummy module is
present into the free site where a detector module is not installed. See
Figure 525. Add a Front/Back/Aux L/Aux R Detector Module
Dummy Front/ Back
Detector Module
J13
on
TRACE 1600/1610 GC
AUX DET LEFT
AUX DET RIGHT
page
518.
Figure
525.
Dummy AUX L/AUX R
Detector Module
TRACE 1600 Auxiliary Oven